 | |  |  Failure Modes in Medical Device Software An Analysis of Fifteen Years of Recall Data
 By D. Richard Kuhn/Dolores R. Wallace

 
 Summary: This paper presents an analysis of software-related failures of medical devices that caused no death or injury but led to recalls by the manufacturers. The analysis categorizes the failures by their symptoms and faults, and discusses methods of preventing and detecting faults in each category. The nature of the faults provides lessons about the value of generally accepted quality practices for prevention and detection methods applied prior to system release. It also provides some insight into the need for formal requirements specification and for improved testing of complex hardware-software systems. |  |  |

|
|
View Content Detail: Failure Modes In Medical Device Software.pdf (121 Kb)
About the Author D. Richard Kuhn is a computer scientist in the Computer Security Division of the National Institute of Standards and Technology (NIST). He is a senior member of the Institute of Electrical and Electronics Engineers (IEEE), received a US Department of Commerce gold medal for scientific/engineering achievement in 2002 and 1998 Excellence in Technology Transfer award from the Federal Laboratory Consortium for research in role based access control, a US Department of Commerce bronze medal in 1990 for contributions to open system standards, and is a member of Beta Gamma Sigma. From 1994 to 1995, he served as Program Manager for the Committee on Applications and Technology of the President’s Information Infrastructure Task Force and from 1996 to 1999 as manager of the Software Quality Group at NIST. He received an MS in computer science from the University of Maryland at College Park, and an MBA from the College of William and Mary in Virginia.
Back to Top
|